Command reference
R&S
®
ZNA
991User Manual 1178.6462.02 ─ 29
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<DC Gain> *RST: 0 dB
Default unit: dB
Options: R&S ZNA-K20
Manual operation: See "CTLE Equalizer" on page 820
CALCulate<Chn>:EYE:EQUalization:CTLE:POLE<1|2> <CTLE Poles>
Specifies the poles of the CTLE (a two-pole filter with single zero) used at the receiver
simulation of the related eye diagram.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
The equalizer can be enabled/disabled using CALCulate<Chn>:EYE:
EQUalization:STATe.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<1|2>
Parameters:
<CTLE Poles> Default unit: Hz
Options: R&S ZNA-K20
Manual operation: See "CTLE Equalizer" on page 820
CALCulate<Chn>:EYE:EQUalization:CTLE:ZERO <CTLE Zero>
Specifies the zero of the equalizer (a two-pole filter with single zero) used at the
receiver simulation of the related eye diagram.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
The equalizer can be enabled/disabled using CALCulate<Chn>:EYE:
EQUalization:STATe.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<CTLE Zero> Default unit: Hz
Options: R&S ZNA-K20
Manual operation: See "CTLE Equalizer" on page 820
SCPI command reference