Command reference
R&S
®
ZNA
994User Manual 1178.6462.02 ─ 29
Manual operation: See "Length" on page 810
CALCulate<Chn>:EYE:INPut:MODulation <PAMType>
Defines or queries the modulation type of the signal to be simulated for the related eye
diagram.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<PAMType> NRZ | PM4 | PM8 | PM16
NRZ, PAM-4, PAM-8 or PAM-16
*RST: NRZ
Manual operation: See "Modulation" on page 811
CALCulate<Chn>:EYE:INPut:XRATe <Rate>
Defines the transfer rate of the bit stream generator for the related eye diagram.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Rate> Transfer rate with units XPS (= transfers per second), KXPS
(=10^3 XPS), MXPS (=10^6 XPS), GXPS (=10^9 XPS)
Default unit: XPS
Options: R&S ZNA-K20
Manual operation: See "Symbol Rate/Transfer Rate" on page 810
CALCulate<Chn>:EYE:INPut:OLEVel <OneLevel>
CALCulate<Chn>:EYE:INPut:ZLEVel <VoltageLevel>
Defines the highest/lowest (nominal) voltage level of the simulated digital signal gener-
ating the related eye diagram.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<VoltageLevel> Default unit: V
Options: R&S ZNA-K20
Manual operation: See "High Level / Low Level" on page 811
SCPI command reference