Command reference
R&S
®
ZNA
1053User Manual 1178.6462.02 ─ 29
Suffix:
<Chn>
.
Channel number used to identify the active trace. If unspecified
the numeric suffix is set to 1.
Setting parameters:
<Activate> ONCE
Initiates the calculation
Example:
*RST; CALC:FORM PHAS
Reset the analyzer; assign the phase of the transmission param-
eter S21 to the default trace.
CALC:LDEV:AUTO ONCE
Calculate correction factors and apply them to the default trace.
CALC:LDEV:SLOP?; ELEN?; CONS?
Query the values of the correction factors.
CALC:LDEV:MODE OFF
Disable the correction.
Usage: Setting only
Manual operation: See "Auto Linearity Deviation" on page 476
CALCulate<Chn>:LDEViation:CONStant <Constant>
Defines the constant value for the linearity deviation calculation.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Constant> Constant value
Range: The range depends on the trace format. For the
default trace: -200 dB to + 200 dB.
Increment: 10E-2 dB
*RST: 0 dB
Example: See CALCulate<Chn>:LDEViation:AUTO on page 1052
Manual operation: See "Constant" on page 477
CALCulate<Chn>:LDEViation:ELENgth <ElLength>
Defines the electrical length for the linearity deviation calculation. This command is
available if the active trace format is Phase or Unwrapped Phase, or if a mixer delay is
measured.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
SCPI command reference