Command reference
R&S
®
ZNA
1054User Manual 1178.6462.02 ─ 29
Parameters:
<ElLength> Electrical length
Range: 0 m to + 1000 m.
Increment: 10E-4 m
*RST: 0 m
Default unit: m
Example: See CALCulate<Chn>:LDEViation:AUTO on page 1052
Manual operation: See "El. Length" on page 477
CALCulate<Chn>:LDEViation:MODE
<Mode>
Applies /discards the correction factors or re-calculates them for each trace (Tracking).
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Mode> OFF | ON | TRACking
Apply the correction using the latest correction factors.
ON - Apply the correction using the latest correction factors.
OFF - Discard the correction.
TRACking - Repeat calculation for each measured trace and
apply the correction.
*RST: 0 dB/MHz
Example: See CALCulate<Chn>:LDEViation:AUTO on page 1052
Manual operation: See "On" on page 476
CALCulate<Chn>:LDEViation:SLOPe <Slope>
Defines the slope of the regression line for the linearity deviation calculation.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Slope> Slope of the regression line
Range: The range depends on the trace format. For the
default trace: -1015 dB/MHz to +1015 dB/MHz.
Increment: 10E-4 dB/MHz
*RST: 0 dB/MHz
Example: See CALCulate<Chn>:LDEViation:AUTO on page 1052
Manual operation: See "Slope" on page 477
SCPI command reference