EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 1056

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1056User Manual 1178.6462.02 ─ 29
Suffix:
<Chn>
.
Channel number used to identify the active trace
Usage: Event
Manual operation: See "Clear Test" on page 495
CALCulate<Chn>:LIMit:CIRCle:DATA <CenterX>, <CenterY>, <Radius>
Defines a circle limit lines by its center coordinates and its radius.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<CenterX> Range: Virtually no restriction for center coordinates.
*RST: 0
Default unit: NN
<CenterY> Range: Virtually no restriction for center coordinates.
*RST: 0
Default unit: NN
<Radius> Range: Virtually no restriction for radius (use positive val-
ues).
*RST: 1
Default unit: NN
Example: See CALCulate<Chn>:LIMit:CIRCle[:STATe]
Manual operation: See "Radius / Center X / Center Y" on page 495
CALCulate<Chn>:LIMit:CIRCle:DISPlay[:STATe] <Boolean>
Displays or hides the circle limit line associated to the active trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF - Circle limit line on or off.
*RST: OFF
Example:
*RST; CALCulate:LIMit:CIRCle:DATA 0, 0, 0.5
Define a circle limit line centered around the origin of the polar
diagram, assigning a radius of 0.5 U.
CALCulate:FORMat POLar
CALCulate:LIMit:CIRCle:DISPlay ON
Activate a polar diagram and show the circle limit line in the dia-
gram.
Manual operation: See "Show Limit Circle" on page 492
SCPI command reference

Table of Contents

Related product manuals