GUI reference
R&S
®
ZNA
495User Manual 1178.6462.02 ─ 29
Figure 5-16: Simultaneous dB Mag limit line and circle check
Remote command:
CALCulate<Chn>:LIMit:CIRCle[:STATe]
CALCulate<Chn>:LIMit:CIRCle:FAIL?
CALCulate:LIMit:CIRCle:FAIL:ALL?
Limit Fail Beep
Activates or deactivates the fail beep. The fail beep is a low-tone acoustic signal that is
generated each time the analyzer detects an exceeded limit. No fail beep can be gen-
erated if the limit check is switched off.
Note: The R&S ZNA does not have a built-in audio device and loudspeaker. To hear
sounds, connect a USB audio device to the instrument or operate it via remote desk-
top.
Remote command:
CALCulate<Chn>:LIMit:CIRCle:SOUNd[:STATe]
Clear Test
Resets the limit check results.
Remote command:
CALCulate<Chn>:LIMit:CIRCle:CLEar
Draw Circle
Activates touchscreen or mouse operation; tap the diagram at one border of the limit
circle and draw the circle to the required size and position.
Remote command:
n/a
Radius / Center X / Center Y
Defines the limit circle by its radius and its center on the X-axis and Y-axis.