Command reference
R&S
®
ZNA
1126User Manual 1178.6462.02 ─ 29
Parameters:
<Marker> Marker number
If marker <Marker> is OFF for the related trace, the set com-
mand is silently ignored.
Manual operation: See "Selected Marker" on page 460
CALCulate<Chn>:STATistics:NLINear:COMP:RPHase <Phase>
Manually defines the reference phase for compression point calculation on phase for-
matted traces.
Only applies if CALCulate<Chn>:STATistics:NLINear:COMP:REFerence is set
to DVALue.
For dB formatted traces, the reference level can be set using CALCulate<Chn>:
STATistics:NLINear:COMP:RLEVel.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Phase> Reference phase
*RST:
Default unit: deg
Manual operation: See "Defined Value" on page 461
CALCulate<Chn>:STATistics:NLINear:COMP[:STATe] <Boolean>
Displays or hides the compression point result in the diagram area channel <Chn>'s
active trace..
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF - statistical info field on or off.
*RST: OFF
Example: See CALCulate<Chn>:STATistics:NLINear:COMP:
RESult?
Manual operation: See "Compr. Point / Compr. Val." on page 459
CALCulate<Chn>:STATistics:PRATio[:STATe] <Boolean>
Displays or hides the power ratio statistics in the diagram area of spectrum analysis
channel <Chn>'s active trace.
Use CALCulate<Chn>:STATistics:PRATio:RESult? to query the results.
SCPI command reference