Command reference
R&S
®
ZNA
1141User Manual 1178.6462.02 ─ 29
Manual operation:  See "Time Start / Time Stop / Time Center / Time Span"
on page 534
Note: If the x-axis is scaled in distance units (CALCulate<Chn>:TRANsform:TIME:
XAXis DISTance), then the span is entered in m; the range and default value
changes accordingly.
CALCulate<Chn>:TRANsform:TIME:STARt <StartTime>
Defines the start time of the diagram in time domain.
Suffix: 
<Chn>
.
Channel number used to identify the active trace
Parameters:
<StartTime> Start time of the diagram.
Range:  -100 s to +99.999999999998 s.
Increment:  0.1 ns
*RST:  -1E-009 s
Default unit: s
Example: 
*RST; :CALC:TRAN:TIME:STAT ON
Reset the instrument, activating a frequency sweep, and enable
the time domain transformation for the default trace.
CALC:TRAN:TIME:STAR 0; STOP 10 ns
Set the start time to 0 ns and the stop time to 10 ns.
Manual operation:  See "Time Start / Time Stop / Time Center / Time Span"
on page 534
Note: If the x-axis is scaled in distance units (CALCulate<Chn>:TRANsform:TIME:
XAXis DISTance), then the start value is entered in m; the range and default value
changes accordingly.
CALCulate<Chn>:TRANsform:TIME:STATe <Boolean>
Determines whether the time domain transformation for trace no. <Chn> is enabled.
Suffix: 
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON - time domain representation active.
OFF - frequency domain representation active.
*RST:  OFF
Example: 
*RST; :CALC:TRAN:TIME:STAT?
Reset the instrument, activating a frequency sweep, and query
whether the default trace is displayed in the time domain. The
response is 0.
Manual operation:  See "Time Domain" on page 448
SCPI command reference