Command reference
R&S
®
ZNA
1142User Manual 1178.6462.02 ─ 29
CALCulate<Chn>:TRANsform:TIME:STIMulus <Type>
Selects the type of stimulus to be simulated in the low pass transformation process.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Type> IMPulse | STEP
IMPulse - impulse response, in bandpass or lowpass mode.
STEP - step response, only in lowpass mode (a bandpass mode
setting CALCulate<Chn>:TRANsform:TIME[:TYPE] BPASs
is automatically changed to lowpass).
*RST: IMP
Example: See CALCulate<Chn>:TRANsform:TIME[:TYPE]
Manual operation: See "Type" on page 449
CALCulate<Chn>:TRANsform:TIME:STOP
<StopTime>
Defines the stop time of the diagram in time domain.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<StopTime> Stop time of the diagram.
Range: -99.999999999998 s to +100 s.
Increment: 0.1 ns
*RST: +4E-009 s
Default unit: s
Example: See CALCulate<Chn>:TRANsform:TIME:STARt
Manual operation: See "Time Start / Time Stop / Time Center / Time Span"
on page 534
Note: If the x-axis is scaled in distance units (CALCulate<Chn>:TRANsform:TIME:
XAXis DISTance), then the stop value is entered in m; the range and default value
changes accordingly.
CALCulate<Chn>:TRANsform:TIME[:TYPE] <TransformType>
Selects the time domain transformation type.
Suffix:
<Chn>
.
Channel number used to identify the active trace
SCPI command reference