Suffix:
<Ch>
.
Channel number
<ListId> Index of the affected port pair (see CALCulate<Ch>:
TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
DEFine)
<1|2|3> Index i of the capacitance C<i> in the related lumped circuit
model. If unspecified the numeric suffix is set to 1.
Parameters:
<Capacitance> Capacitance Ci
Range: -1 mF to 1 mF
Increment: 1 fF (1E-15 F)
*RST: 1 pF (1E-12 F)
Default unit: F
Parameters for setting and query:
<CircuitModel> STSC | SCST | CSSL | LSSC | CSSC | SLCS | SCLS | SCCS |
STSG | SGST | GSSL | LSSG | GSSG | SLGS | SGLS | SGGS
Circuit model whose capacitance C<i> shall be set, see
Table 7-7
Example: See CALCulate<Ch>:TRANsform:VNETworks:PPAir:
DEEMbedding<ListId>[:STATe]
Manual operation: See "Network" on page 743
CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
PARameters:G<1|2|3> <CircuitModel>, <Conductance>
CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
PARameters:G<1|2|3>? <CircuitModel>
Specifies the conductance value G<i> in the different lumped circuit models for port
pair deembedding.
In the query form, the <Conductance> parameter must be omitted. The command
returns the conductance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<ListId> Index of the affected port pair (see CALCulate<Ch>:
TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
DEFine)
<1|2|3> Index i of the conductance G<i> in the related lumped circuit
model. If unspecified the numeric suffix is set to 1.