Command reference
R&S
®
ZNA
1200User Manual 1178.6462.02 ─ 29
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Mode> MAGPhase | REIMag
Selects the averaging mode.
MAGPhase
Averaging of magnitude and phase (default) of the complex
trace value
REIMag
Averaging of real and imaginary part of the complex trace value
Manual operation: See "Mode" on page 466
CALCulate<Chn>:IAVerage[:STATe] <Boolean>
Turns infinite averaging on or off.
The averaging mode can be selected using CALCulate<Chn>:IAVerage:MODE.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF | 1 | 0
*RST: 0
Manual operation: See "Reset History" on page 399
CALCulate<Chn>:PHOLd <HoldFunc>
Enables, disables, or restarts the max hold and the min hold functions.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<HoldFunc> MIN | MAX | OFF
MIN - Enable the min hold function.
MAX - enable the max hold function.
OFF - disable the max hold or min hold function.
*RST: OFF
Example:
*RST; :CALC:PHOL MAX
Reset the instrument and enable the max hold function.
CALC:PHOL OFF; PHOL MAX
Restart max hold.
Manual operation: See "Reset History" on page 399
SCPI command reference