Command reference
R&S
®
ZNA
1201User Manual 1178.6462.02 ─ 29
CALCulate<Chn>:SMOothing:APERture <SmoothAperture>
Defines how many measurement points are averaged to smooth the trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<SmoothAperture> Smoothing aperture. A smoothing aperture of n % means that
the smoothing interval for each sweep point i with stimulus value
x
i
is equal to [x
i
- span*n/200, x
i
+ span*n/200], and that the
result of i is replaced by the arithmetic mean value of all mea-
surement points in this interval.
Range: 0.05% to 100%.
*RST: 1
Default unit: %
Example:
*RST; :CALC:SMO ON
Activate smoothing for the default trace.
CALC:SMO:APER 0.5
Reduce the smoothing aperture to 0.5 %.
Manual operation: See "Aperture" on page 464
CALCulate<Chn>:SMOothing[:STATe] <Boolean>
Enables or disables smoothing for trace no. <Chn>.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF - smoothing on or off.
*RST: OFF
Example: See CALCulate<Chn>:SMOothing:APERture
Manual operation: See "Smoothing" on page 463
CALCulate<Ch>:TDIF:IMBalance:COMPensation[:STATe] <Boolean>
Selects the calculation method for S-parameters, ratios and derived quantities during
an amplitude imbalance or phase imbalance sweep.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
ON
Compensate imbalance of a waves (see "Imbalance Compensa-
tion of a-Waves" on page 278)
SCPI command reference