Command reference
R&S
®
ZNA
1219User Manual 1178.6462.02 ─ 29
CONTrol<Ch>:RFFE<Bus>:SETTings:FREQuency <ClockFrequency>
Sets/gets the clock frequency for channel <Ch> and RFFE bus <Bus>.
Suffix:
<Ch>
.
Channel number
<Bus> 1 or 2
RFFE bus number
Parameters:
<ClockFrequency> Clock rate.
Possible values are 52/1664 MHz, 52/1663 MHz, ..., 52/2 MHz.
Range: 31.25 kHz to 26 MHz
Default unit: Hz
Manual operation: See "CLK, VIO, VLow, VHigh" on page 660
CONTrol<Ch>:RFFE<Bus>:SETTings[:STATe] <EnableInSequence>
Enables/disables RFFE bus interface <Bus> in the Sweep Sequencer for channel
<Ch> (see CONTrol<Ch>:SEQuence<Nr>:RFFE<Bus>:COMMand:DATA and
CONTrol<Ch>:SEGMent<SegNr>:SEQuence<SeqNr>:RFFE<Bus>:COMMand:
DATA).
Suffix:
<Ch>
.
Channel number
<Bus> RFFE bus interface number
Parameters:
<EnableInSequence> Enabled state
Manual operation: See "Seq." on page 659
CONTrol<Ch>:RFFE<Bus>:SETTings:VOLTage:HIGH <VoltageHigh>
CONTrol<Ch>:RFFE<Bus>:SETTings:VOLTage:IO <IOVoltage>
CONTrol<Ch>:RFFE<Bus>:SETTings:VOLTage:LOW <Voltage>
Sets/gets the IO, Low and High voltages for channel <Ch> and RFFE bus interface
<Bus>, respectively.
Note that these voltages are only applied while an RFFE command is executed using
CONTrol<Ch>:RFFE<Bus>:COMMand:SEND or CONTrol<Ch>:RFFE<Bus>:
COMMand:SEND?.
Suffix:
<Ch>
.
Channel number
<Bus> RFFE bus number
Parameters:
<Voltage> Range: 0 to 2.5 V
Increment: 0.001 V
Default unit: V
SCPI command reference