Command reference
R&S
®
ZNA
1225User Manual 1178.6462.02 ─ 29
<SegNr> Segment number
<SeqNr> Sequence number, defining the order in which the commands
shall be executed.
For every channel, segment and RFFE interface, the sequence
numbers must be consecutive, starting at 1; reusing a sequence
number overwrites a previously defined command.
<Bus> RFFE bus interface number
Parameters:
<Command> 3 to 37 hexadecimal digits (0-F), defining the command to be
executed: digit 1 is the slave address, digits 2 and 3 specify the
command number and the remaining digits represent the data
part (up to 17 digit pairs).
Manual operation: See "RFFE columns (sweep sequencer table)" on page 666
CONTrol<Ch>:SEGMent<Nr>:SEQuence:CLEar:ALL
For segmented sweeps this command deletes the command/switch sequence for the
respective channel and sweep segment (defined using CONTrol<Ch>:
SEGMent<SegNr>:SEQuence<SeqNr>:RFFE<Bus>:COMMand:DATA and
CONTrol<Ch>:SEGMent<SegNr>:SEQuence<SeqNr>:GPIO<Port>:VOLTage).
Suffix:
<Ch>
.
Channel number
<Nr> Segment number
Usage: Event
Manual operation: See "Range" on page 666
CONTrol<Ch>:SEGMent<Nr>:SEQuence:COUNt?
For segmented sweeps this command queries the length of the command/switch
sequence defined for a particular channel and sweep segment.
The command/switch sequences are defined using CONTrol<Ch>:
SEGMent<SegNr>:SEQuence<SeqNr>:RFFE<Bus>:COMMand:DATA and
CONTrol<Ch>:SEGMent<SegNr>:SEQuence<SeqNr>:GPIO<Port>:VOLTage.
Suffix:
<Ch>
.
Channel number
<Nr> Segment number
Usage: Query only
Manual operation: See "Range" on page 666