Manual operation: See "Trace Info" on page 846
DISPlay:ANNotation:FREQuency[:STATe] <Boolean>
Unmasks or masks all stimulus values in the VNA GUI for the active recall set.
Parameters:
<Boolean> ON | OFF - unmask or mask stimulus values
*RST: ON - all stimulus values unmasked
Example:
*RST; :DISP:ANN:FREQ OFF
Create diagram area no. 1 (with default trace) and mask all stim-
ulus values.
Manual operation: See "Hide Sensitive Information" on page 846
DISPlay:CMAP:LIMit:FCOLorize[:STATe] <Boolean>
Assigns a different trace color to failed trace segments ("Colorize Trace when Failed").
Parameters:
<Boolean> ON | OFF - colorize trace or keep original trace color.
*RST: n/a (a *RST does not affect the setting). In the fac-
tory configuration, OFF is preset.
Example: See DISPlay:CMAP:LIMit[:STATe]
Manual operation: See "Limit Test > Colorize Trace when Failed" on page 849
DISPlay:CMAP:LIMit:FSYMbol[:STATe] <Boolean>
Displays or hides the limit fail symbols (colored squares) on the trace.
Parameters:
<Boolean> ON | OFF - show or hide symbols.
*RST: n/a (a *RST does not affect the setting). In the fac-
tory configuration, ON is preset.
Example: See DISPlay:CMAP:LIMit[:STATe]
Manual operation: See "Limit Test > Show Limit Fail Symbols" on page 848
DISPlay:CMAP:LIMit[:STATe] <Boolean>
Displays all limit lines either with individually configured colors or with the color of the
associated trace(s). The colors of all display elements are defined via DISPlay:
CMAP<DispEl>:RGB.
.
SCPI command reference

Table of Contents