Command reference
R&S
®
ZNA
1396User Manual 1178.6462.02 ─ 29
[SENSe<Ch>:]CORRection:METas:CABLe:MOVement [<Ports>[, <Ports>]...]
While a METAS calibration is active for the related channel (see [SENSe<Ch>:
]CORRection:METas[:STATe] on page 1397), this command notifies the METAS
VNA Tools II about additional cable movements at the specified ports.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<Ports> <integer>
First port number, where cables were moved
ALL
Cable movements at all calibrated ports
NONE
Clears the cable movement history in the METAS project
<Ports> Additional port numbers, where cables were moved
Usage: Setting only
Options: R&S ZNA-K50
Manual operation: See "Add New Connection/Add Cable Movement" on page 652
[SENSe<Ch>:]CORRection:METas:CABLe:REConnection [<Ports>[, <Ports>]...]
While a METAS calibration is active for the related channel (see [SENSe<Ch>:
]CORRection:METas[:STATe] on page 1397), this command notifies the METAS
VNA Tools II about additional connector reconnections at the specified ports.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<Ports> <integer>
First port number, where connectors were reconnected
ALL
All ports were reconnected
NONE
Clears the connector reconnection history in the METAS project
<Ports> Additional port numbers, where connectors were reconnected
Usage: Setting only
Options: R&S ZNA-K50
Manual operation: See "Add New Connection/Add Cable Movement" on page 652
[SENSe<Ch>:]CORRection:METas:CABLe:TYPE <CableNameFile>[, <Ports>[,
<Ports>]...]
Selects the METAS cable type that is used at the respective ports.
SCPI command reference