Command reference
R&S
®
ZNA
1397User Manual 1178.6462.02 ─ 29
Use [SENSe:]CORRection:METas:CABLe:CATalog? to list the available cable
types.
Suffix:
<Ch>
.
Channel number
The related channel must be prepared for a METAS calibration
using [SENSe<Ch>:]CORRection:COLLect:METHod:
DEFine<CalName>, MTOSm | MUOSm | MOSM |
MTRL, ....
Setting parameters:
<CableNameFile> Name of a METAS cable model file (without *.cable exten-
sion) in the database of the local METAS VNA Tools II installa-
tion.
<Ports> First port number or ALL, if the same cable type is used for all
ports.
The latter is equivalent to omitting all <Ports> parameters.
<Ports> Additional port numbers
Usage: Setting only
Options: R&S ZNA-K50
Manual operation: See "Cable" on page 646
[SENSe<Ch>:]CORRection:METas[:STATe] <Boolean>
After a successful METAS calibration, this command activates or deactivates the
uncertainty calculation for the related channel.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1)
METAS calibration active
OFF (0)
METAS calibration inactive
Options: R&S ZNA-K50
Manual operation: See "METAS Cal Active" on page 642
[SENSe<Ch>:]CORRection:METas:UNCertainty <Level>
Defines the width of the uncertainty band, expressed as multiples of the standard devi-
ation.
Default is 2 (95% confidence interval).
Suffix:
<Ch>
.
Channel number
SCPI command reference