Command reference
R&S
®
ZNA
1405User Manual 1178.6462.02 ─ 29
TYPE
Calibration type (see <Type> parameter in [SENSe<Ch>:
]CORRection:COLLect[:ACQuire]:SELected
on page 1356)
PORTs
The related test port numbers (comma-separated list of integers)
SPORt
For SMARTer calibrations this returns the port to which a power-
meter was connected (0 otherwise)
THRoughs
Measued Throughs: comma-separated list of ordered port num-
ber pairs <pno1-pno2>
TSTamp
Timestamp (in UTC)
LTSTamp
Timestamp in local time
TVNA
For calibrations involving switch matrices, this indicates the test
ports on the VNA itself. Returns a comma-separated list of port
pairs TestPort,VnaPort
MVNA
For calibrations involving switch matrices, this indicates the
mapping between matrix ports and VNA ports for matrix
<Index>. Returns a comma-separated list of port pairs
MatrixVnaPort,VnaPort
MTESt
For calibrations involving switch matrices, this indicates the
mapping between matrix (physical) test ports and test ports for
matrix <Index>. Returns a comma-separated list of port pairs
MatrixTestPort,TestPort
CKIT
For each port, the name of the used calibration kit is stored with
the calibration.
The query can return one of the following:
the cal kit name (string)
if available and unique
'Multiple'
if cal kit names are available but not unique, i.e. if <PhyPt> was
not specified and multiple cal kits were used (multi-port cals with
different connector types, merged cals)
<Index>
Currently not used