Command reference
R&S
®
ZNA
1439User Manual 1178.6462.02 ─ 29
Example: See [SENSe<Ch>:]FREQuency:IMODulation:LTONe
[SENSe<Ch>:]FREQuency:IMODulation:RECeiver <PortNumber>
Selects the receiving port for the intermodulation measurement.
Suffix:
<Ch>
.
Channel number
Parameters:
<PortNumber> Analyzer port number
Range: 1 to the number of analyzer ports
*RST: 2
Example: See [SENSe<Ch>:]FREQuency:IMODulation:LTONe
Manual operation: See "Receiving Port" on page 386
[SENSe<Ch>:]FREQuency:IMODulation:SPECtrum:MORDer <IMOrder>
Defines the maximum order of intermodulation products for the intermodulation spec-
trum measurement.
Suffix:
<Ch>
.
Channel number
Parameters:
<IMOrder> Maximum order of IM products
Range: 3 | 5 | 7 | 9
*RST: 3
Example: See [SENSe<Ch>:]FREQuency:IMODulation:SPECtrum[:
STATe]
Manual operation: See "CW Mode Spectrum" on page 381
[SENSe<Ch>:]FREQuency:IMODulation:SPECtrum[:STATe] <Boolean>
Enables or disables the measurement of the intermodulation spectrum. If a new chan-
nel is desired, the trace must be created separately; see second example below.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON | OFF - enables or disables intermodulation spectrum mea-
surement.
*RST: OFF