Command reference
R&S
®
ZNA
1483User Manual 1178.6462.02 ─ 29
<Unused> Ignored parameter (for compatibility with R&S ZVR analyzers).
Should be set to the default value 0.
<MeasBandwidth> IF bandwidth in the segment. See [SENSe<Ch>:
]SEGMent<Seg>:BWIDth[:RESolution].
Default unit: Hz
<LO> POSitive | NEGative | AUTO
Position of the local oscillator frequency LO relative to the RF
frequency. In remote control this parameter must be set when a
sweep segment is created.
See [SENSe<Ch>:]FREQuency:SBANd.
<Mode> AUTO
Ignored.
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings.
SEGM:DEF?
Query the channel settings for the new segment. A possible
response is
1000000000,2000000000,51,-300,0.0044625,0,
10000,AUTO,NORM,AUTO,STEP.
Manual operation: See "Optional Columns" on page 557
[SENSe<Ch>:]SEGMent<Seg>:DEFine:SELect <TimeRef>
Defines whether the sweep time of a new segment, i.e. numeric parameter no. 5 of the
command [SENSe<Ch>:]SEGMent<Seg>:DEFine, is entered as a segment sweep
time ("Segment Time") or as a measurement delay ("Meas Delay").
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number
Parameters:
<TimeRef> SWTime | DWELl
SWTime - use segment sweep time.
DWELl - use measurement delay.
Example:
SEGM1:DEF:SEL DWEL
Select the measurement delay to determine the sweep time in a
new sweep segment no. 1.
SEGM1:DEF 1MHZ, 1.5MHZ, 111, -21DBM, 0.01S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz and a measurement delay of 10 ms.
SEGM1:SWE:TIME?
Query the sweep time in the new segment.
SCPI command reference