1479User Manual 1178.6462.02 ─ 29
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz.
SEGM2:ADD
Create a second sweep segment. The frequency range of the
second segment will be between 1.5 MHz and the maximum fre-
quency of the analyzer.
Usage: Event
Manual operation: See "Add / Insert / Delete / Delete All" on page 554
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution] <ResBandwidth>
Defines the IF bandwidth of the analyzer (measurement bandwidth) in sweep segment
no. <Seg>. Values between 1 Hz and 1.5 MHz can be set. Option R&S ZNA-K17 ena-
bles IF bandwidths up to 10 MHz (see Chapter 4.7.8, "Increased IF bandwidth
30 MHz", on page 288).
Bandwidths can be set in 1 1.5 2 3 5 7 steps. The analyzer rounds up any
entered value between these steps and rounds down values exceeding the maximum
bandwidth.
At the same time, the command activates separate bandwidth setting in all sweep seg-
ments ([SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:CONTrol ON).
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number
Parameters:
<ResBandwidth> IF bandwidth
Range: See above
Increment: 1-1.5-2-3-5-7 steps
*RST: 10 kHz
Default unit: Hz
Example: See [SENSe<Ch>:]SEGMent<Seg>:BWIDth[:
RESolution]:CONTrol
Manual operation: See "Optional Columns" on page 557
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:CONTrol <Boolean>
Selects common or independent "Meas. Bandwidth" settings for the sweep segments.
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number. This suffix is ignored; the setting con-
trols the whole segmented sweep.