Command reference
R&S
®
ZNA
1485User Manual 1178.6462.02 ─ 29
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz.
SEGM:FREQ:CENT? SPAN?
Query the center frequency and span of the created segment.
The response is 1250000;500000.
Usage: Query only
Manual operation: See "Table Columns" on page 554
Note: The frequency range of the sweep segment can be changed via [SENSe<Ch>:
]SEGMent<Seg>:FREQuency:STARt and [SENSe<Ch>:]SEGMent<Seg>:
FREQuency:STOP.
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt <StartFreq>
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP <StopFreq>
These commands define the start and stop frequency of sweep segment no. <Seg>.
The sweep segments must be within the frequency range of the R&S ZNA; see
Table 7-17.
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number
Parameters:
<StopFreq> Start or stop frequency of the sweep.
Increment: 0.1 kHz
Default unit: Hz
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz.
SEGM:FREQ:STAR?; STOP?
Query the start and stop frequency of the created segment. The
response is 1000000;1500000.
Manual operation: See "Table Columns" on page 554
Note: If the start frequency entered is greater than the current stop frequency, the stop
frequency is set to the start frequency. If the stop frequency entered is smaller than the
current start frequency, the start frequency is set to the stop frequency.
[SENSe<Ch>:]SEGMent<Seg>:INSert <StartFreq>, <StopFreq>, <Points>, <Power>,
<SegmentTime>|<MeasDelay>, <Unused>, <MeasBandwidth>[, <LO>,
<Selectivity>, <Mode>, <FreqSweepMode>]
Adds a new sweep segment no. <Seg> with specific channel settings.
SCPI command reference