Command reference
R&S
®
ZNA
1539User Manual 1178.6462.02 ─ 29
Note that this only takes effect if SOURce<Ch>:GROup<Grp>:SIMultaneous:
FOFFset:MOFFset:MODE is set to DIRect.
Suffix:
<Ch>
.
Channel number.
If unspecified, this parameter is set to 1.
<Grp>
Parameters:
<StartFreq> The minimum frequency offset.
To get the resulting frequency offset, the minimum frequency off-
set is rounded to a multiple of the current frequency step size.
Range: 0 Hz to 17 MHz
*RST: 1 MHz
Default unit: Hz
Example: See SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:
STATe
Manual operation: See "Minimum Offset" on page 685
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:MOFFset:MODE <arg0>
This command is used to define how the minimum frequency offset between different
port groups is determined for Parallel Measurement with Frequency Offset.
The minimum frequency offset specifies the minimum frequency difference between
the measurements performed in parallel in any two different port groups. The fre-
quency offset which is actually used may be larger than the minimum offset as the firm-
ware chooses the smallest multiple of the frequency step size between the points in a
segment or in a linear frequency sweep, respectively, which is larger than the minimum
offset.
Suffix:
<Ch>
.
Channel number.
If unspecified, this parameter is set to 1.
<Grp>
Parameters:
<arg0> DIRect | BANDwidth
DIRect
The minimum frequency offset to be used is directly specified by
the command SOURce<Ch>:GROup<Grp>:SIMultaneous:
FOFFset:MOFFset:DVALue.
SCPI command reference