Command reference
R&S
®
ZNA
1562User Manual 1178.6462.02 ─ 29
SOURce:POWer:CORRection:TCOefficient:FEED <Trace>[, <DeEmbedding
Position>]
Selects a trace which provides the points for the transmission coefficient list.
Parameters:
<DeEmbedding
Position>
BOTH | DUT | PM
Determines the two-port the command refers to:
DUT
The two-port between VNA and DUT (during measurement)
PM
The two-port between VNA and PM (during power calibration)
BOTH
Both positions
(parameter omitted)
If the parameter is omitted, the command refers to the two-port
selected using SOURce<Ch>:POWer<PhyPt>:CORRection:
TCOefficient:CALibration.
Setting parameters:
<Trace> String parameter for the trace name, e.g. 'Trc1'. The trace must
exist in the active recall set.
*RST: n/a
Example: See SOURce:POWer:CORRection:TCOefficient[:STATe]
on page 1563
Usage: Setting only
Manual operation: See "Get Trace..." on page 634
SOURce:POWer:CORRection:TCOefficient:INSert<ListNo> <Frequency>,
<TransCoeff>[, <DeEmbedding Position>]
Inserts a new frequency and transmission coefficient at position no. <ListNo> in the
transmission coefficient list. The following points are shifted down in the list.
Suffix:
<ListNo>
.
List entry no.
Parameters:
<Frequency> Stimulus frequency value. If several points with identical fre-
quencies are added, the analyzer automatically ensures a fre-
quency spacing of 1 Hz.
Range: Stimulus values outside the frequency range of the
analyzer are allowed.
*RST: n/a
Default unit: Hz
SCPI command reference