SOURce<Ch>:POWer<Pt>:EATTenuation:MODE <ElSourceAttMode>
Defines how the electronic source attenuation of port <Pt> is selected in channel
<Ch>.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<ElSourceAttMode> AUTO | MANual
AUTO
The firmware selects the suitable electronic attenuation value.
MANual
You can select the electronic attenuation value using
SOURce<Ch>:POWer<Pt>:EATTenuation.
Manual operation: See "El. Source Att." on page 676
SOURce<Ch>:POWer<PhyPt>:GENerator<Gen>:OFFSet <Offset>, <Mode>
Defines the power of an external generator or its power offset relative to the channel
power (SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate][:AMPLitude]).
Suffix:
<Ch>
.
Channel number
<PhyPt> Test port number of the analyzer. This suffix is ignored; the gen-
erator is selected via <Gen>.
<Gen> Generator number
Parameters:
<Offset> Port-specific power offset
Range: -300 dB to +300 dB (adjust to the generator's actual
source power range and the test setup)
Increment: 0.01 dB (other values are rounded)
*RST: 0 dB
Default unit: dB
<Mode> ONLY | CPADd
ONLY - only the port-specific power is used; the channel power
is ignored.
CPADd - the port-specific power is added as an offset to the
channel power.
*RST: CPADd