Command reference
R&S
®
ZNA
1591User Manual 1178.6462.02 ─ 29
Suffix:
<Ch>
.
Channel number
<PhyPt> Test port number of the analyzer
Parameters:
<Boolean> ON - power at port <PhyPt> is permanently on.
OFF - power is only on for the partial measurements that require
the port as a drive port.
*RST: OFF
Example:
SOUR:POW2:STAT ON
Switch the RF power at port no. 2 on.
SOUR:POW2:PERM ON
Activate a permanent signal source.
Manual operation: See "Source Gen" on page 671
SOURce:POWer:REDuce:SDELay <SettlingDelay>
Deprecated; superseded by SOURce:POWer:SWEepend:SDELay
If power reduction at sweep end is enabled (SOURce<Ch>:POWer<PhyPt>:
REDuce[:STATe] ON) this defines the settling delay at subsequent sweep starts, i.e.
the time between power-up and sweep start when a new sweep is requested.
This is a global setting.
Parameters:
<SettlingDelay> Settling delay
Default unit: s
Example: See SOURce:POWer:SWEepend:MODE on page 1593
SOURce<Ch>:POWer<PhyPt>:REDuce[:STATe] <ReduceAtSweepEnd>
Deprecated; superseded by SOURce:POWer:SWEepend:MODE
Enables/disables power reduction at sweep end.
This is a global setting.
Suffix:
<Ch>
.
Channel number. This suffix is ignored; the setting is valid for all
channels.
<PhyPt> Physical port number. This suffix is ignored; the setting is valid
for all ports.
Parameters:
<ReduceAt
SweepEnd>
OFF (default): at sweep end the output power of the sweep start
is restored
ON: at sweep end the output power of the driving port is reduced
as if the channel base power was set to its minimum possible
value
SCPI command reference