Command reference
R&S
®
ZNA
1611User Manual 1178.6462.02 ─ 29
Example:
SYST:COMM:RDEV:AKAL:ADDR:ALL?
Query the names of all connected calibration units.
SYST:COMM:RDEV:AKAL:ADDR 'ZN-Z51::1234'
Select the cal unit named 'ZN-Z51::1234'.
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:CATalog?
Query all characterizations stored on the connected calibration
unit.
SYSTem:COMMunicate:RDEVice:AKAL:SDATa?
'My_calkit', OPEN, S11, 1
Query the characterization data for the characterization named
'My_calkit' and an open standard (one-port, port restriction). A
characterization with the queried properties must be available on
the cal unit.
SYSTem:COMMunicate:RDEVice:AKAL:WARMup?
Query the warmup status of the calibration unit.
SYSTem:COMMunicate:RDEVice:AKAL:DATE? 'Factory'
Query the creation date of the factory calibration.
SYSTem:COMMunicate:RDEVice:AKAL:FRANge?
'Factory'
Query the frequency range of the factory calibration.
SYSTem:COMMunicate:RDEVice:AKAL:PORTs?
'Factory'
Query the port assignment of the factory calibration.
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:STANdard:
CATalog? 'Factory'
Query the standards of the factory calibration. Possible
response:
'MOP(P1),MSH(P1),MMMT(P1),MOP(P2),MSH(P2),
MMMT(P2)' denotes an Open (m), Short (m) and Match (m)
standard at each of the ports 1 and 2.
Usage:
Query only
Manual operation: See "Characterization" on page 587
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:STANdard:CATalog? <CalKitName>
Queries all calibration standards of a characterization (calkit file) that is stored on a
connected calibration unit, either on the calibration unit's internal (flash) memory or on
an SD card inserted at the calibration unit (if available). The query SYSTem:
COMMunicate:RDEVice:AKAL:CKIT:CATalog? returns the names of the cal kit
files. The factory characterization is always available.
Query parameters:
<CalKitName> Name of the calkit file, string parameter.
'Factory' denotes the factory characterization, the prefix
"SD:" indicates that the characterization is stored on an SD card
inserted at the calibration unit.