Command reference
R&S
®
ZNA
1636User Manual 1178.6462.02 ─ 29
Parameters:
<Boolean> ON | OFF - enable/disable display of information popups for
remote control errors
*RST: ON
Example:
SYST:ERR:DISP:STATe ON; REM ON
Switch the display of information popups for remote command
errors on.
FREQ:STAR -1
Generate a Remote Error: -222, "Data out of
range; ..." tooltip.
Manual operation: See "Show Remote Error Info Messages" on page 864
SYSTem:ERRor:DISPlay:STATe <Boolean>
Globally defines whether instrument events shall be indicated by information popups.
Display of popups can be limited to certain event types using commands SYSTem:
ERRor:DISPlay:INFO, SYSTem:ERRor:DISPlay:WARNings, SYSTem:ERRor:
DISPlay:ERRor, and SYSTem:ERRor:DISPlay[:REMote].
Parameters:
<Boolean> ON | OFF - globally enable or disable the display of information
popups
*RST: ON
Example: See SYSTem:ERRor:DISPlay:ERRor and SYSTem:ERRor:
DISPlay[:REMote].
Manual operation: See "Show Instrument Messages" on page 864
SYSTem:ERRor[:NEXT]?
Queries and at the same time deletes the oldest entry in the error queue. Operation is
identical to that of STATus:QUEue[:NEXT]?
The entry consists of an error number and a short description of the error. Positive
error numbers are instrument-dependent. Negative error numbers are reserved by the
SCPI standard; see section Error messages and troubleshooting.
Example:
SYST:ERR?
Query the oldest entry in the error queue. 0,"No error" is
returned if the error queue is empty.
Usage: Query only
SYSTem:FIRMware:UPDate <VNASetupFile>
Installs a firmware version stored in a VNA setup file (*.exe) on the analyzer. The
installation is automatic and does not require any further action.
SCPI command reference