Command reference
R&S
®
ZNA
1640User Manual 1178.6462.02 ─ 29
Also sets:
The (default) byte order for binary data transfer (see FORMat:BORDer)
The (default) identification (*IDN?) and option (*OPT?) strings (see SYSTem:
IDENtify:FACTory and SYSTem:OPTions:FACTory).
Parameters:
<Language> 'SCPI'
R&S ZNA-specific command set: the analyzer supports all com-
mands described in this documentation.
Factory default. The byte order is set to SWAPped.
'ZVABT' | 'ZVR'
Compatibility mode for network analyzers of the R&S ZVR and
R&S ZVA/B/T families.
The byte order is set to SWAPped.
'ENA' | 'E5071' | 'PNA' | 'HP8510' | 'HP8530' | 'HP8714' |
'HP8720' | 'HP8753'
Compatibility modes for network analyzers from other manufac-
turers.
The byte order is set to NORMal.
Example:
SYST:LANG 'PNA'
Select a PNA-compatible command set.
Manual operation: See "Remote Language" on page 888
SYSTem:LOGGing:REMote[:STATe] <Boolean>
Enables logging of all remote control commands transferred to the analyzer.
Parameters:
<Boolean> ON command sequence stored to file
C:\Users\Public\Documents\Rohde-Schwarz\ZNA\
RemoteLog\VNARemote.log.
OFF command sequence not logged.
Example:
SYST:LOGG:REM ON
Enable remote logging. The log file
'C:\Users\Public\Documents\Rohde-Schwarz\ZNA\
RemoteLog\VNARemote.log' is fixed and does not have to
be specified.
SYSTem:OPCHeck:PLEVel:PORT<Port>? <Power>[, <LimitLines>]
Defines and executes a power level operator check on port <Port> and returns its boo-
lean result.
Suffix:
<Port>
.
Physical VNA port no.
SCPI command reference