Command reference
R&S
®
ZNA
1639User Manual 1178.6462.02 ─ 29
SYSTem:HELP:SYNTax:ALL?
Queries the implemented SCPI commands and their parameters.
Return values:
<Commands>
Usage: Query only
SYSTem:IDENtify:FACTory
Resets the response to the *IDN? query to the factory ID, overwriting a user-defined
identification string (SYSTem:IDENtify).
Use SYSTem:FORMat:IDENtify to change the format of the factory ID string.
Example:
SYSTem:IDENtify:STRing 'MyDevice'; *IDN?
Define an identity string. The response is MyDevice.
SYSTem:IDENtify:FACTory; *IDN?
Re-activate the factory setting.
Usage: Event
Manual operation: See "Define *IDN + *OPT..." on page 889
SYSTem:IDENtify[:STRing] <InstId>
Defines an identity string for the network analyzer. The query is equivalent to *IDN?.
Parameters:
<InstId> String parameter containing the instrument identity
Example: See SYSTem:IDENtify:FACTory
Manual operation: See "Define *IDN + *OPT..." on page 889
SYSTem:KLOCk <Boolean>
Locks or unlocks the local controls of the analyzer. This includes the front panel keys,
the keyboard, or other local interfaces.
Parameters:
<Boolean> ON | OFF - lock or unlock the local keys.
*RST: OFF
Example:
SYST:KLOCk ON
Lock the local keys.
SYSTem:LANGuage <Language>
Defines the remote language of the analyzer, i.e. which commands are available and
how they are interpreted.
SCPI command reference