Command reference
R&S
®
ZNA
1654User Manual 1178.6462.02 ─ 29
Suffix:
<Ch>
.
Channel number.
Parameters:
<DelayTime> Delay time.
Range: 0 s to 13680 s
Increment: 10 ms
*RST: 0 s
Default unit: s
Example:
TRIG:SOUR MAN
Activate external trigger source.
TRIG:HOLD UP
Set a delay time of 10 ms.
Manual operation: See "Delay" on page 562
TRIGger<Ch>[:SEQuence]:LINK <MeasSequence>
Selects the triggered measurement sequence. The identifier for the sequence is a
string variable.
Suffix:
<Ch>
.
Channel number
Parameters:
<MeasSequence> Triggered measurement sequence, string variable.
'SWEep' trigger event starts an entire sweep.
'SEGMent' trigger event starts a sweep segment, if segmented
frequency sweep is active (see example below). If another
sweep type is active, the trigger event starts an entire sweep.
'POINt' trigger event starts measurement at the next sweep
point.
'PPOint' trigger event starts the next partial measurement at
the current or at the next sweep point.
*RST: 'SWEep'
Example:
SEGM:ADD; :SWE:TYPE SEGM
Select segmented frequency sweep.
TRIG:LINK 'SEGM'
Select a trigger segment as triggered measurement sequence.
TRIG:LINK?
Query the triggered measurement sequence. The response is
'SEGMENT'.
Manual operation: See "Sequence" on page 561