EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 1655

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1655User Manual 1178.6462.02 ─ 29
TRIGger[:SEQuence]:LTRigger[:STATe] <Boolean>
Activates/deactivates tracking of lost trigger events.
Parameters:
<Boolean> ON (1)
Lost trigger traclking active
OFF (0)
Lost trigger tracking inactive
Manual operation: See "Activate Lost Trigger" on page 880
TRIGger[:SEQuence]:LTRigger:COUNt?
Returns the number of lost trigger events the analyzer firmware observed since the last
reset of the lost trigger event log (see TRIGger[:SEQuence]:LTRigger:RESet
on page 1655).
Usage:
Query only
Manual operation: See "Sequence" on page 561
TRIGger[:SEQuence]:LTRigger:RESet
Clears the lost trigger event log.
Usage:
Event
Manual operation: See "Clear Lost Trigger Log" on page 880
TRIGger<Ch>[:SEQuence]:MULTiple:HOLDoff <MeasSequence>[, <DelayTime>]
Sets a delay time between the trigger event and the start of the measurement ("Trigger
Delay") in multiple trigger mode.
Suffix:
<Ch>
.
Channel number
Parameters:
<MeasSequence> SWEep | SEGMent | POINt | PPOint
Triggered measurement sequence, PPOint denotes "partial
measurement"; see TRIGger<Ch>[:SEQuence]:LINK.
<DelayTime> Delay time
Range: 0 s to 13680 s
Increment: 10 ms
*RST: 0 s
Default unit: s
Example: See TRIGger<Ch>[:SEQuence]:MULTiple:SOURce<inp>
Manual operation: See "... /Logic/ ..." on page 563
SCPI command reference

Table of Contents

Related product manuals