EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 1688

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1688User Manual 1178.6462.02 ─ 29
Usage: Query only
DIAGnostic:HUMS:UTILization:ACTivity:TRACking:STATe <ID>, <State>
DIAGnostic:HUMS:UTILization:ACTivity:TRACking:STATe? <ID>
Turns the activity tracking of utilizations on and off.
Parameters:
<State> ON | OFF | 1 | 0
ON | 1
Tracking activity = on
OFF | 0
Tracking activity = off
Parameters for setting and query:
<ID> ID number of the utilization activity.
To find out the ID of the utilization, you have to query the com-
plete utilization data. For more information, see DIAGnostic:
HUMS:UTILization? on page 1687.
Example: //Set activity tracking state of ID 31719936
DIAG:HUMS:UTIL:ACT:TRAC:STAT 31719936,'1'
DIAGnostic:HUMS:UTILization:CUSTom:ADD <ID>, <Name>, <Description>,
<Unit>, <State>
Creates a custom utilization.
Setting parameters:
<ID> Identifier of the custom utilization.
Range: 1 to 99
<Name> Name of the custom utilization that appears in the user interface.
<Description> Description of the custom utilization.
<Unit> Unit of the custom utilization that appears in the user interface.
<State> ON | OFF | 1 | 0
Initial activity state of custom utilization.
HUMS and service date commands

Table of Contents

Related product manuals