GUI reference
R&S
®
ZNA
410User Manual 1178.6462.02 ─ 29
In its upper part, the dialog shows a diagram with the RF and LO signals, the mixing
product (IF), and the current frequency and power ranges for all signals. The diagram
is adjusted according to the "2nd Mixer" selection; see "Two-stage mixer measure-
ments" on page 271.
The lower part of the dialog allows you to configure the mixer stages and to define
sweep, stimulus and other channel settings.
Port selection
The port lists in the graphical part of the dialog contain all analyzer ports or external
generators which can provide the RF signal and local oscillator (LO) signals.
Note: If External switch matrices are used, operation with more than one internal
source RF is not supported. In this case, only the Converter LO port or an external
generator can be used as local oscillator.
Generators must be configured explicitly in the External Generators dialog before they
appear in the list. "None" means that the input signal at LO (if available) is not control-
led by the analyzer.
Remote command:
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:RFPort
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:IFPort
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:LOPort<Stg>
Multiplication factors
The fractional numbers in the graphical part of the dialog indicate the frequency con-
version factors for the RF and LO signals.