Command reference
R&S
®
ZNA
1432User Manual 1178.6462.02 ─ 29
Parameters:
<Numerator>
Range: 1, 2, 3 ...
*RST: 1
<Denominator> Range: 1, 2, 3 ...
*RST: 1
Example: See [SENSe<Ch>:]FREQuency:CONVersion:MIXer:
MFFixed
Manual operation: See "Multiplication factors" on page 410
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:LOPort<Stg> <SourceType>[,
<SourceNumber>]
Selects an analyzer or an external generator port as a signal source for the LO1 or
LO2 signal.
Note:
If External switch matrices are part of the RF connection configuration, operation with
more than one internal source is not supported. In this case only external generators
can be used as local oscillators.
Suffix:
<Ch>
.
Channel number
<Stg> Mixer stage (signals LO 1 or LO 2; see [SENSe<Ch>:
]FREQuency:CONVersion:MIXer:STAGes).
Parameters:
<SourceType> PORT | GENerator | NONE
NONE LO signal is not controlled by the analyzer.
PORT analyzer port
GENerator external generator
*RST: NONE
<SourceNumber> Number of analyzer port or generator.
Example: See [SENSe<Ch>:]FREQuency:CONVersion:MIXer:
MFFixed
Manual operation: See "Port selection" on page 410
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:MFFixed <InputOutput>[,
<FixedFrequency>]
Assigns a fixed frequency to the RF, LO 1, LO 2, or to the IF signal. The fixed fre-
quency setting becomes active if the port is selected as a port with fixed frequency
([SENSe<Ch>:]FREQuency:CONVersion:MIXer:FIXed<Stg>).