GUI reference
R&S
®
ZNA
807User Manual 1178.6462.02 ─ 29
Advanced Settings...
If the active trace is an eye diagram, the "Advanced Settings..." button opens the
Advanced Settings dialog that allows in-depth configuration of the simulated digital sig-
nal generator and receiver.
Display Measurements
If the active trace is an eye diagram, the "Display Measurements" button in the Eye
Diag tab softtool tab toggles the display of the eye diagram results.
The available results depend on the selected Modulation:
●
For NRZ modulated generator signals, the result display comprises two separate
info fields for "Basic" and "Time" results (see "Measurements..." on page 807). By
default they are stacked below each other, however, they can be moved independ-
ently like any other info field.
Figure 5-60: Eye Diagram Results (NRZ Modulation)
●
For PAM modulation types, only a reduced set of "Basic" results is available.
Figure 5-61: Eye Diagram Results (PAM Modulations)
For a description of the result values, see "Eye Diagram Results" on page 257.
Remote command:
CALCulate<Chn>:EYE:MEASurement:STATe
CALCulate<Chn>:EYE:MEASurement:DATA?
Measurements...
If the active trace is an eye diagram, the "Measurements..." button brings up a dialog
which allows to select and configure the eye measurements to be displayed (if Display
Measurements is enabled).
Applic softtool