GUI reference
R&S
®
ZNA
817User Manual 1178.6462.02 ─ 29
Active
This button must be first checked to activate any of the jitter sources. At this point,
each of the individual jitter sources can be activated.
Remote command:
CALCulate<Chn>:EYE:JITTer:STATe
Random
Random jitter is determined from a normal (Gaussian) random distribution whose
mean is zero, and whose "Std. Deviation" in seconds is specified by the user.
Remote command:
CALCulate<Chn>:EYE:JITTer:TYPE:RANDom
CALCulate<Chn>:EYE:JITTer:RANDom:STDDeviation
Periodic
This type of jitter is determined by a sine wave whose amplitude ("Periodic Magni-
tude"), frequency ("Periodic Frequency") and phase ("Periodic Phase") is specified by
the user.
Remote command:
CALCulate<Chn>:EYE:JITTer:TYPE:PERiodic
CALCulate<Chn>:EYE:JITTer:PERiodic:MAGNitude
CALCulate<Chn>:EYE:JITTer:PERiodic:FREQuency
CALCulate<Chn>:EYE:JITTer:PERiodic:PHASe
Dirac
This type of jitter is specified by the amplitude in seconds ("Dirac Delta", positive or
negative), and the probability of the jitter occurring at each symbol period ("Dirac Prob-
ability").
Remote command:
CALCulate<Chn>:EYE:JITTer:TYPE:DIRac
CALCulate<Chn>:EYE:JITTer:DIRac:DELTa
CALCulate<Chn>:EYE:JITTer:DIRac:PROBability
User Specific
The "Load Jitter" button allows the user to open a "Jitter File (*.txt)" with each entry in
the file equal to the jitter to be added at each symbol period.
Applic softtool