Command reference
R&S
®
ZNA
1018User Manual 1178.6462.02 ─ 29
The deembedding assistant functionality requires at least one of the fixture deembed-
ding software options:
●
R&S ZNA-K220 (see Chapter 4.7.16, "In-situ de-embedding", on page 299)
●
R&S ZNA-K230 (see Chapter 4.7.17, "Smart fixture de-embedding", on page 300)
●
R&S ZNA-K210 (see Chapter 4.7.15, "Eazy de-embedding based on IEEE 370",
on page 298)
If options are of these available, the CALCulate:FMODel:DEASsistant commands
cannot be used.
CALCulate:FMODel:DEASsistant:COUPon:LEFT:CLEar
CALCulate:FMODel:DEASsistant:COUPon:RIGHt:CLEar
Clears the test coupon data available for the lead-in/lead-out.
Usage:
Event
Options: R&S ZNA-K220 | R&S ZNA-K230 | R&S ZNA-K210
Manual operation: See "Coupon A / Coupon B section" on page 761
CALCulate:FMODel:DEASsistant:COUPon:LEFT:FILE <CouponType>, <Path>
CALCulate:FMODel:DEASsistant:COUPon:RIGHt:FILE <CouponType>, <Path>
Loads the properties of a test coupon for the lead-in/lead-out of the DUT.
Parameters:
<CouponType> SYMMetric2x | OPEN1x | SHORt1x
<Path> Path to the Touchstone file, either absolute or relative to the
selected working directory (CALCulate:FMODel:DIRectory).
For
CALCulate:FMODel:DEASsistant:COUPon:
<LEFT|RIGHt>:TYPE OPSHort1x, you have to load a file for
OPEN1x and SHORt1x (or measure it). Otherwise only for the
selected coupon type.
Options: R&S ZNA-K220 | R&S ZNA-K230 | R&S ZNA-K210
Manual operation: See "Deembedding Files..." on page 756
CALCulate:FMODel:DEASsistant:COUPon:LEFT:MEASure <CouponType>
CALCulate:FMODel:DEASsistant:COUPon:RIGHt:MEASure <CouponType>
Measures the test coupon.
Before measuring, select the test coupon type using CALCulate:FMODel:DEAS
sistant:COUPon:<LEFT|RIGHt>:TYPE and the ports using CALCulate:FMO
Del:DEASsistant:COUPon:<LEFT|RIGHt>:PORT.
SCPI command reference