Setting parameters:
<CouponType> SYMMetric2x | OPEN1x | SHORt1x
Test coupon type to be measured
For
CALCulate:FMODel:DEASsistant:COUPon:
<LEFT|RIGHt>:TYPE OPSHort1x, you have to measure
OPEN1x and SHORt1x (or load it from file). Otherwise measure
only the selected coupon type.
Usage: Setting only
Options: R&S ZNA-K220 | R&S ZNA-K230 | R&S ZNA-K210
Manual operation: See "Coupon A / Coupon B section" on page 761
CALCulate:FMODel:DEASsistant:COUPon:LEFT:PORT <Position>, <PortNum>
CALCulate:FMODel:DEASsistant:COUPon:RIGHt:PORT <Position>, <PortNum>
Defines the port or ports for measuring the selected test coupon (CALCulate:
FMODel:DEASsistant:COUPon:RIGHt:TYPE)
Parameters:
<Position> UPLeft | UPRight
UPLeft
Left side of the test coupon
UPRight
Right side of the test coupon (SYMMetric2x only)
Setting parameters:
<PortNum> Port number (physical port number for single-ended ports, logi-
cal port number for balanced ports)
Options: R&S ZNA-K220 | R&S ZNA-K230 | R&S ZNA-K210
Manual operation: See "Coupon A / Coupon B section" on page 761
CALCulate:FMODel:DEASsistant:COUPon:LEFT:TYPE <CouponType>
CALCulate:FMODel:DEASsistant:COUPon:RIGHt:TYPE <CouponType>
For fixture modeling with the deembedding assistant, this command selects a test cou-
pon to be measured for the lead-in/lead-out, or loaded from file.
The valid test coupons depend on the selected deembedding tool (CALCulate:
FMODel:DEASsistant:TOOL).
SYMMetric2x OPEN1x SHORt1x OPSHort1x
ISD