EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 1020

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1020User Manual 1178.6462.02 ─ 29
CALCulate:FMODel:DEASsistant:COUPon:<LEFT|RIGHt>:PORT to
select the port(s) at which the test coupon is measured.
CALCulate:FMODel:DEASsistant:COUPon:<LEFT|RIGHt>:MEASure
To load the measurement data from file, use CALCulate:FMODel:DEASsist
ant:COUPon:<LEFT|RIGHt>:FILE.
Parameters:
<CouponType> SYMMetric2x | OPEN1x | SHORt1x | OPSHort1x
Test coupon type
SYMMetric2x
Symmetric 2 x Thru
OPEN1x
1 x Open
SHORt1x
1 x Short
Options: R&S ZNA-K220 | R&S ZNA-K230 | R&S ZNA-K210
Manual operation: See "Coupon A / Coupon B section" on page 761
CALCulate:FMODel:DEASsistant:DUT:MEASure
Measures DUT + fixture at the ports defined using CALCulate:FMODel:
DEASsistant:DUT:PORT.
Usage:
Event
Options: R&S ZNA-K220 | R&S ZNA-K230 | R&S ZNA-K210
Manual operation: See "Measure / Apply section" on page 763
CALCulate:FMODel:DEASsistant:DUT:PORT <Position>, <PortNum>
Before you can measure DUT + fixture or run the fixture deembedding tool
(CALCulate:FMODel:DEASsistant:RUN), you have to define consecutive port
ranges for the left/right side of the fixture.
Parameters:
<Position> UPLeft | UPRight | LOLeft | LORight
Defines the port ranges.
For DUTs with one single-ended or balanced ports on the left/
right, specifying an up-port is sufficient for this side.
UPLeft
Upper end of the port range on the left
UPRight
Upper end of the port range on the right
LOLeft
Lower end of the port range on the left
SCPI command reference

Table of Contents

Related product manuals