Command reference
R&S
®
ZNA
1022User Manual 1178.6462.02 ─ 29
CALCulate:FMODel:DEASsistant<Ph_pt>:FIXTure:LEFT:TYPE <FixtureType>
CALCulate:FMODel:DEASsistant:FIXTure:RIGHt:TYPE <FixtureType>
Allows you to specify models for the left side (lead-in) and right side (lead-out) of the
DUT, depending on the selected DUT type (CALCulate:FMODel:DEASsistant:
DUT:TYPE on page 1021).
Parameters:
<FixtureType> NONE | SE | BALanced | COUPled | UNCoupled
NONE
Only for non-existing right sides (lead-outs).
SE
Single-ended; for single-ended lead-ins/outs
BALanced
Balanced; for balanced lead-ins/outs
COUPled
Coupled ports; for lead-ins/outs with more than two logical ports
UNCoupled
Coupled ports; for lead-ins/outs with more than two physical
ports
Options:
R&S ZNA-K220 | R&S ZNA-K230 | R&S ZNA-K210
Manual operation: See "Left: Model A / Right: Model A|B" on page 758
CALCulate:FMODel:DEASsistant:IMPCorrect <Boolean>
Parameters:
<Boolean>
Manual operation: See "Measure / Apply section" on page 763
CALCulate:FMODel:DEASsistant:PRESet
Resets the DUT and fixture settings of the deembedding assistant to their respective
defaults.
Note that a *RST does not affect these settings.
Usage:
Event
Options: R&S ZNA-K220 | R&S ZNA-K230 | R&S ZNA-K210
Manual operation: See "DUT" on page 757
CALCulate:FMODel:DEASsistant:RUN
Runs the selected demmbedding (CALCulate:FMODel:DEASsistant:TOOL) tool
and applies the calculated fixture deembedding to the ports selected using
CALCulate:FMODel:DEASsistant:DUT:PORT.
Usage:
Event