Command reference
R&S
®
ZNA
1027User Manual 1178.6462.02 ─ 29
CALCulate:FMODel:DELT<Ph_pt>:FREQuencies:USER:ADD|DELete adds/
deletes a frequency to|from the list of custom frequencies.
Setting parameters:
<Frequency> Custom frequency
Range: The frequency range of the analyzer
Default unit: Hz
Usage: Setting only
Options: R&S ZNA-K231
Manual operation: See "Frequencies" on page 766
CALCulate:FMODel:DELT:FREQuencies:USER:DELete:ALL
The Delta-L algorithm can calculate the loss-per-inch uncertainties either at default or
at custom frequencies (see CALCulate:FMODel:DELT:FREQuencies:
USEDefault).
This command clears the list of custom frequencies.
Usage:
Event
Options: R&S ZNA-K231
Manual operation: See "Frequencies" on page 766
CALCulate:FMODel:DELT:FREQuencies:USER:COUNt?
The Delta-L algorithm can calculate the loss-per-inch uncertainties either at default or
at custom frequencies (see CALCulate:FMODel:DELT:FREQuencies:
USEDefault).
This command returns the number of custom frequencies, defined using CALCulate:
FMODel:DELT:FREQuencies:USER:ADD etc.
Usage:
Query only
Options: R&S ZNA-K231
Manual operation: See "Frequencies" on page 766
CALCulate:FMODel:DELT:M1L:CACHe:CLEar:ALL
If the 1-length method is used (CALCulate:FMODel:DELT:METHod M1L), the VNA
firmware builds up a cache of coupon measurements (Touchstone files) in the
DeltaL\1LMeasurementCache subfolder of the user data folder
C:\Users\Public\Documents\Rohde-Schwarz\ZNA.
This command clears this cache.
Usage:
Event
Options: R&S ZNA-K231
SCPI command reference