Command reference
R&S
®
ZNA
1028User Manual 1178.6462.02 ─ 29
Manual operation: See "Clear All 1L Measurements/Clear Selected 1L Measure-
ments" on page 769
CALCulate:FMODel:DELT:M1L:CACHe:CLEar:SELected <String>
If the 1-length method is used (CALCulate:FMODel:DELT:METHod M1L), the VNA
firmware builds up a cache of coupon measurements (Touchstone files) in the
DeltaL\1LMeasurementCache subfolder of the user data folder
C:\Users\Public\Documents\Rohde-Schwarz\ZNA.
This command deletes the selected file from this cache.
Setting parameters:
<String> Path to the Touchstone file, relative to the cache directory.
Usage: Setting only
Options: R&S ZNA-K231
Manual operation: See "Clear All 1L Measurements/Clear Selected 1L Measure-
ments" on page 769
CALCulate:FMODel:DELT:M1L:DIFFmode <Boolean>
The one-length method (CALCulate:FMODel:DELT:METHod M1L) handles single-
ended and balanced measurement results differently. This command selects between
the two modes.
Parameters:
<Boolean> OFF (0)
Single-ended mode
ON (1)
Differential mode
*RST: OFF (0)
Options: R&S ZNA-K231
Manual operation: See "Method" on page 765
CALCulate:FMODel:DELT:M1L:FILename <String>
CALCulate:FMODel:DELT:M2L:FILename <String>
CALCulate:FMODel:DELT:M3L:FILename <String>
For a Delta-L+ PCB characterization, this command loads the "1L Measurement" data
from a Touchstone file.
Parameters:
<String> Path to the Touchstone file, either absolute or relative to the
selected working directory (CALCulate:FMODel:DELT:
DIRectory).
Options: R&S ZNA-K231
SCPI command reference