Command reference
R&S
®
ZNA
1029User Manual 1178.6462.02 ─ 29
Manual operation: See "Load File" on page 768
CALCulate:FMODel:DELT:M1L:LENGth <Length>
CALCulate:FMODel:DELT:M2L:LENGth <Length>
CALCulate:FMODel:DELT:M3L:LENGth <Length>
Specifies the length of the 1st/2nd/3rd line in Delta-L measurements (CALCu
late:FMODel:DELT<Ph_pt>:M1L|M2L|M3L:MEASure).
Parameters:
<Length> Length in inches
Options: R&S ZNA-K231
Manual operation: See "Length" on page 769
CALCulate:FMODel:DELT:M1L:MEASure
CALCulate:FMODel:DELT:M2L:MEASure
CALCulate:FMODel:DELT:M3L:MEASure
Measures the full set of S-parameters at the active ports (see CALCulate:FMO
Del:DELT<Ph_pt>:M1L|M2L|M3L[:STATe]).
Usage:
Event
Options: R&S ZNA-K231
Manual operation: See "Measure" on page 769
CALCulate:FMODel:DELT<Ph_pt>:M1L[:STATe] <Boolean>
CALCulate:FMODel:DELT<Ph_pt>:M2L[:STATe] <Boolean>
CALCulate:FMODel:DELT<Ph_pt>:M3L[:STATe] <Boolean>
Defines the active ports for the 1st/2nd/3rd line Delta-L measurement (CALCu
late:FMODel:DELT<Ph_pt>:M1L|M2L|M3L:MEASure).
For each measurement, either two or four ports must be active. If the 1-length method
is used (CALCulate:FMODel:DELT:METHod M1L), then four ports must be active for
the first line.
Suffix:
<Ph_pt>
.
Physical port number
Parameters:
<Boolean> ON (1)
Port <Ph_pt> is active in the respective measurement
OFF (0)
Port <Ph_pt> is inactive in the respective measurement
Options: R&S ZNA-K231
Manual operation: See "Ports table" on page 769