Command reference
R&S
®
ZNA
1042User Manual 1178.6462.02 ─ 29
CALCulate:FMODel:ISD:DUT:MEASure
For a fixture modeling with the ISD tool, this command allows to measure DUT + Fix-
ture at the active ports (see CALCulate:FMODel:ISD<Ph_pt>:DUT[:STATe])
Example: See Chapter 8.2.9, "Fixture modeling", on page 1762
Usage: Event
Manual operation: See "Measure" on page 733
CALCulate:FMODel:ISD:DUT:MEASure:FILename
<String>
For a fixture modeling with the ISD tool, this command loads the "DUT + Test Fixture"
properties from a Touchstone file.
Parameters:
<String> Path to the Touchstone file, either absolute or relative to the
selected working directory (CALCulate:FMODel:DIRectory).
Manual operation: See "Load File" on page 733
CALCulate:FMODel:ISD:DUT:MEASure:FILename:CLEar
Invalidates the file path previously set using CALCulate:FMODel:ISD:DUT:
MEASure:FILename.
Usage:
Event
Manual operation: See "Load File" on page 733
CALCulate:FMODel:ISD<Ph_pt>:DUT[:STATe] <Boolean>
For a fixture modeling with the ISD tool, this command allows to specify the ports to
which the test fixture is connected.
This has to be defined before measuring DUT + Fixture (using CALCulate:FMODel:
ISD:DUT:MEASure).
Suffix:
<Ph_pt>
.
Physical port number
Parameters:
<Boolean> 1 (ON, true) if the test fixture is connected to port <Ph_pt>, 0
(OFF, false) otherwise
Example: See Chapter 8.2.9, "Fixture modeling", on page 1762
Manual operation: See "Active" on page 733
CALCulate:FMODel:ISD:DUT:TYPE <DUTType>
Defines whether the DUT is active or passive.
SCPI command reference