Command reference
R&S
®
ZNA
1129User Manual 1178.6462.02 ─ 29
ELENgth - return the electrical length.
PDELay - return the phase delay.
GAIN - return the gain, i.e. the larger of two marker values.
SLOPe - return the slope (difference) between two marker val-
ues.
FLATness - return the flatness of the trace between two marker
positions.
ALL - return all statistical values, observing the order used
above.
The data is returned as a comma-separated list of real numbers.
The unit is the default unit of the measured parameter (see
CALCulate<Ch>:PARameter:SDEFine) but may also depend
on the trace format (see CALCulate<Chn>:FORMat). For com-
plex traces the statistical results MEAN, STDDev, MAX, MIN,
RMS and PTPeak are calculated in the selected format (see
CALCulate<Chn>:STATistics:FORMat).
Example:
*RST; :CALC:STAT:RES? MAX
Calculate and return the maximum of the default trace showing
an S-parameter on a dB Mag scale.
:CALC:FORM POL; STAT:RES? MAX
Display the trace in a polar diagram and re-calculate the maxi-
mum. The result corresponds to the previous result but is con-
verted to a unitless linear value.
Usage: Query only
Manual operation: See "Min/Max/Peak-Peak, Mean/Std Dev/RMS" on page 456
CALCulate<Chn>:STATistics:RMS[:STATe] <Boolean>
CALCulate<Chn>:STATistics:SFLatness[:STATe] <Boolean>
These commands display or hide the "RMS" and the "Flatness/Gain/Slope" results in
the diagram area of trace no. <Chn>.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF - statistical info field on or off.
*RST: OFF
Example: See CALCulate<Chn>:STATistics[:STATe]
Manual operation: See "Flatness/Gain/Slope" on page 458
CALCulate<Chn>:STATistics[:STATe] <Boolean>
Displays or hides all statistical results in the diagram area of trace no. <Chn> except
the compression point results.
Tip: You can display or hide the "Min/Max/Peak-Peak", "Mean/Std Dev/RMS",
"Phase/El Length" and "Flatness/Gain/Slope" results separately; see example below.
SCPI command reference