Command reference
R&S
®
ZNA
1138User Manual 1178.6462.02 ─ 29
Manual operation: See "Side Lobe Level" on page 449
CALCulate<Chn>:TRANsform:TIME:LPASs <Algorithm>
Calculates the harmonic grid for low pass time domain transforms according to one of
the three alternative algorithms.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Algorithm> KFSTop | KDFRequency | KSDFrequency
KFSTop - keep stop frequency and number of points
KDFRequency - keep frequency gap and number of points
KSDfrequency - keep stop frequency and approximate fre-
quency gap
Example: See CALCulate<Chn>:TRANsform:TIME[:TYPE]
Manual operation: See "Set Harmonic Grid and Keep" on page 451
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam <DCValue>
Defines the DC value for low pass transforms. The command is enabled only if the
sweep points are on a harmonic grid (to be set explicitly or using CALCulate<Chn>:
TRANsform:TIME:LPASs).
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<DCValue> DC value of the measured quantity
Range: Depending on the measured quantity (-1 to +1 for
S-parameters)
*RST: 0
SCPI command reference