Command reference
R&S
®
ZNA
1173User Manual 1178.6462.02 ─ 29
Example:
CALC:TRAN:VNET:GLO:EMB:TND SL
Select the Shunt L circuit model for embedding.
CALC:TRAN:VNET:GLO:EMB:PAR:R SL, 2.2E+3; :CALC:
TRAN:VNET:GLO:EMB ON
Increase the resistance for the Shunt L circuit model to 2.2 kΩ
and enable embedding.
Manual operation: See "Active" on page 714
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>:
TNDefinition <CircuitModel>
Selects the circuit model for ground loop embedding.
Suffix:
<Ch>
.
Channel number.
<group> Port group (DUT) number.
If multiple port groups are configured (see SOURce<Ch>:
GROup<Grp>:PPORts) and CALCulate<Ch>:TRANsform:
VNETworks:GLOop:GROup is ON, then each port group can
have its own de-/embedding models.
Parameters:
<CircuitModel> FIMPort | SL | SC | SG
Possible circuit models (character data); see Circuit models for
ground loop de-/embedding.
*RST: FIMPort
Example: See CALCulate<Ch>:TRANsform:VNETworks:GLOop:
EMBedding<group>[:STATe] on page 1172
Manual operation: See "Network" on page 749
CALCulate<Ch>:TRANsform:VNETworks:GLOop:GROup <Boolean>
If set to ON (1), each port group (defined using SOURce<Ch>:GROup<Grp>:PPORts)
has its own ground loop embedding and deembedding network. Otherwise the same
embedding network and deembedding network is used for all active ports.
See the
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:...
and CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>:...
commands.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> Range: ON (1) | OFF (0)
*RST: OFF (0)
Manual operation: See "Ground Loop per Port Group" on page 751
SCPI command reference

Table of Contents

Related product manuals