GUI reference
R&S
®
ZNA
749User Manual 1178.6462.02 ─ 29
Use the complementary dock widget to activate or deactivate ground loop dembedding
and embedding (see Chapter 5.14.1.6, "Ground Loop panel", on page 713).
If multiple port groups (DUTs) are configured and Ground Loop per Port Group is
enabled, use the dock widget to select the port group whose ground loop network shall
be configured.
Background information
Refer to Chapter 4.6.2.7, "Ground loop de-/embedding", on page 241.
If the "Fixture Simulator" is disabled for the related channel (see "Fixture Simulator"
on page 655), this tab is inactive, i.e. all controls except the "Overview" button are
grayed out.
Overview
See "Overview" on page 717.
Type
Switches between "Deembedding" and "Embedding" network definition.
All Deembedding Activated
Deactivates or activates all deembeddings (single-ended, balanced, port set, ground
loop) for the active channel, if any.
This control is only visible if Type = "Deembedding".
Network
The graphical list contains all available 1-port networks:
The symbol selects "no network" and disables ground loop de-/embedding.
Offset Embed softtool