GUI reference
R&S
®
ZNA
750User Manual 1178.6462.02 ─ 29
The "1-Port Data" network is defined by imported S-parameter data; see Select File...
The remaining networks (Shunt L and Shunt C) are defined by lumped elements
whose parameters are displayed below the graphical list.
Tip:
Drag and drop the network symbols in horizontal or vertical direction to switch to
the next symbol.
Use the Conductance in Embedding Networks switch to change from resistances
in "Capacitor in parallel with resistor" model to conductances and vice versa.
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
TNDefinition
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:C
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:L
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:R
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:G
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>:
TNDefinition
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>:
PARameters:C
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>:
PARameters:L
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>:
PARameters:R
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>:
PARameters:G
Select File...
"Select File..." is enabled as long as the "1-Port Data" network is selected. This net-
work is defined by its S-parameters stored in a one-port Touchstone file (*.s1p). No
additional parameters are required.
Note: The loaded file is stored in the active recall set. Persisted recall sets contain the
full (de-)embedding data so that they can be transferred to other instruments.
Remote command:
MMEMory:LOAD:VNETworks<Ch>:GLOop:DEEMbedding<group>
MMEMory:LOAD:VNETworks<Ch>:GLOop:EMBedding<group>
Set to Ideal Ground
This function is enabled as long as the 1-Port Data network is active. An imported S-
parameter set is replaced by the S-parameters of an ideal through connection, which
eliminates the transformation network.
Offset Embed softtool